OzoneLab Menu / Help
OzoneLab(TM) Instruments
Spacer ImageOzoneLab™ Ozonation Equipment F.A.Q. TopicSetup ExampleProduct PricingHow to Order ProductsSite SearchContact Us
Files ImageHome > Products > Generators

Main characteristics of OzoneLab™ Instruments:

Oxygen/Ozone Gas Flow Range: 0-1000ml/min
Ozone Concentration Range: 0-180mcg/ml (0-12.5 % by weight)
Ozone Output Range: 0-2000mg/hr from single unit 0-4000 or even 0-6000mg/hr from OzoneLab™ Tower systems
Gas Purity Requirements: Laboratory/Medical

If your ozone project falls within above outlined boundaries, it is very likely that you will be able to find an OzoneLab™ instrument that would meet your needs. We encourage you to start your search for the most suitable model right on this page.

OL80 Module (Single Stage) OL100 Module (Double Stage)
OL80 Wall Mounted Line
OL80 Desktop Line
90° Front Panel Enclosure
OL80 Desktop Line - DST-2S Series
90° Front Panel Enclosure
OL80A Ozone Generator & Gas Analyzer
in Desktop Tower Enclosure
OL80M Ozone Generator & Controller
in Desktop Tower Enclosure

Effective September 1st, 2008, an ABS Case with custom cut foam will be included at no additional charge to those clients ordering OL80 Ozone Hardware Set totalling over $3,000 US (excluding S&H).

OL100 Desktop Line
in Tower Enclosure

Typical Ozone Output for single stage
OL80 Ozone Generator

Flow
LPM
Flow
ml/min
Output
µg/ml
Output
[mg/hr]
0 0 0 0
1/32 31 0 - 161 0 - 299
1/16 62 0 - 131 0 - 487
1/8 125 0 - 93 0 - 698
1/4 250 0 - 57 0 - 855
1/2 500 0 - 33 0 - 990
3/4 750 0 - 23 0 - 1035
1 1000 0 - 17 0 - 1020

  • OL80 Sample Output Test Report
  • OL80 Ozone Output Graphs


Creating complex Tower Systems
Typical Ozone Output for double stage
OL100 Ozone Generator

Flow
LPM

Flow
ml/min
Output
µg/ml
Output
[mg/hr]
0 0 0 0
1/32 31 0 - 183 0 - 340
1/16 62 0 - 174 0 - 647
1/8 125 0 - 144 0 - 1080
1/4 250 0 - 100 0 - 1500
1/2 500 0 - 63 0 - 1890
3/4 750 0 - 44 0 - 1980
1 1000 0 - 34 0 - 2040

  • OL100 Sample Output Test Report
  • OL100 Ozone Output Graphs


Creating complex Tower Systems
  • Compare OL80 & OL100
  • Compare OL80, OL100, and GE60 units produced between 1994 & 1999
  • How we test our Instruments
  • Note that we are able to provide Custom Ozone Output adjustments in the range of 10% - 100% of the Typical Ozone Output. Contact us for details.
  • Each ozone generator is tested and calibrated separately using 2" H2O back pressure.
  • OzoneLab™ instruments are designed for low flow/high output applications and are not suitable for flow rates above 1000ml/min due to the size of orifices in the ozone gas flow path. Gas Pressure is 0-3 p.s.i. (For applications dealing with higher pressures - please contact us for more info and/or consultation)

Examples of External Ozone Production & Ozone Flow Control Devices

Foot Switch Control OLW - Transmitter Module C-30Z Ozone Monitor OLS Software
Foot Switch Control  OLW - Transmitter Module C-30Z Ozone Monitor Computer with
OLS OzoneLab™ Software


Discontinued Generators

The following archive links are provided for client information.

  GE30 12VDC GE30 12VDC
Model GE30FM/12VDC GE60FM/12VDC OL80 Desktop Line
45° Front Panel Enclosure
Production Dates: 1994 - 1999 1994 - 1999 2000 - 2007


© Ozone Services For problems pertaining to our Web Site, please contact our WebMaster